Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Sub Yokogawa uR100 (1 year) Voltage CAL VER /5500 DATE: 15-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.4 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 43 NUMBER OF LINES: 163 CONFIGURATION: Fluke 5500A ============================================================================= # # Requires: # Channel # (1-3) in M[1] # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P J F W 1.002 ASK+ K X 1.003 MATH MEM = M[1] 1.004 RSLT =CHANNEL [MEM] VOLTAGE 1.005 HEAD {MEASUREMENT (DISPLAY) & RECORDING ACCURACY} 1.006 DISP Using the keyboard on the side of the UUT... 1.006 DISP 1.006 DISP Set CHANNEL NUMBER to [MEM]. 1.006 DISP Set MEASUREMENT RANGE to 00. 1.006 DISP Set RECORDING RANGE to -20.00mV to 20.00mV. 1.006 DISP 1.006 DISP Refer to instruction manual if unfamiliar w/keyboard 1.007 DISP Using the keyboard on the side of the UUT... 1.007 DISP 1.007 DISP Set the DISPLAY to MAN DATA. 1.007 DISP 1.007 DISP Use the < > keys to select CHANNEL [MEM]. 1.008 DISP The following tests apply an input, test the Display 1.008 DISP Readout, then asks to check the recording pen position 1.008 DISP on the chart, for 0, 50%, 100% and then 50% again. 1.008 DISP 1.008 DISP The second 50% tests that recording returns to the same 1.008 DISP center position from both directions (dead band). 1.009 ASK- R 1.010 DISP Connect the 5500A to UUT CHANNEL [MEM] as follows: 1.010 DISP [32] 5500A NORMAL HI to A - TOP INPUT TERMINAL 1.010 DISP [32] 5500A NORMAL LO to B - BOTTOM INPUT TERMINAL 1.011 5500 00 -20.00mV 0.1% 0.02U 2W 2.001 5500 -20.00mV S 2W 2.002 EVAL Is the CH[MEM] recording at the {left edge of chart +/-1div}? 3.001 5500 00 0.00mV 0.1% 0.02U 2W 4.001 5500 0.00mV S 2W 4.002 EVAL Is the CH[MEM] recording at the {center of chart +/-1div}? 5.001 5500 00 20.00mV 0.1% 0.02U 2W 6.001 5500 20.00mV S 2W 6.002 EVAL Is the CH[MEM] recording at the {right edge of chart +/-1div}? 7.001 5500 0.00mV S 2W 7.002 MESS Is the chart recording at the same center line as the 7.003 EVAL previous center check +/- {1div (dead band)}? 8.001 MESS 8.002 DISP Using the keyboard on the side of the UUT... 8.002 DISP 8.002 DISP Set CHANNEL NUMBER to [MEM]. 8.002 DISP Set MEASUREMENT RANGE to 01. 8.002 DISP Set RECORDING RANGE to -200.0mV to 200.0mV. 8.003 5500 01 -200.0mV 0.1% 0.02U 2W 9.001 5500 -200.0mV S 2W 9.002 EVAL Is the CH[MEM] recording at the {left edge of chart +/-1div}? 10.001 5500 01 0.0mV 0.1% 0.02U 2W 11.001 5500 0.0mV S 2W 11.002 EVAL Is the CH[MEM] recording at the {center of chart +/-1div}? 12.001 5500 01 200.0mV 0.1% 0.02U 2W 13.001 5500 200.0mV S 2W 13.002 EVAL Is the CH[MEM] recording at the {right edge of chart +/-1div}? 14.001 5500 0.0mV S 2W 14.002 MESS Is the chart recording at the same center line as the 14.003 EVAL previous center check +/- {1div (dead band)}? 15.001 MESS 15.002 DISP Using the keyboard on the side of the UUT... 15.002 DISP 15.002 DISP Set the CHANNEL NUMBER to [MEM]. 15.002 DISP Set MEASUREMENT RANGE to 02. 15.002 DISP Set RECORDING RANGE to -2.000V to 2.000V. 15.003 5500 02 -2.000V 0.1% 0.02U 2W 16.001 5500 -2.000V S 2W 16.002 EVAL Is the CH[MEM] recording at the {left edge of chart +/-1div}? 17.001 5500 02 0.000V 0.1% 0.02U 2W 18.001 5500 0.000V S 2W 18.002 EVAL Is the CH[MEM] recording at the {center of chart +/-1div}? 19.001 5500 02 2.000V 0.1% 0.02U 2W 20.001 5500 2.000V S 2W 20.002 EVAL Is the CH[MEM] recording at the {right edge of chart +/-1div}? 21.001 5500 0.000V S 2W 21.002 MESS Is the chart recording at the same center line as the 21.003 EVAL previous center check +/- {1div (dead band)}? 22.001 MESS 22.002 DISP Using the keyboard on the side of the UUT... 22.002 DISP 22.002 DISP Set the CHANNEL NUMBER to [MEM]. 22.002 DISP Set MEASUREMENT RANGE to 03. 22.002 DISP Set RECORDING RANGE to -6.000V to 6.000V. 22.003 5500 03 -6.000V 0.1% 0.02U 2W 23.001 5500 -6.000V S 2W 23.002 EVAL Is the CH[MEM] recording at the {left edge of chart +/-1div}? 24.001 5500 03 0.000V 0.1% 0.02U 2W 25.001 5500 0.000V S 2W 25.002 EVAL Is the CH[MEM] recording at the {center of chart +/-1div}? 26.001 5500 03 6.000V 0.1% 0.02U 2W 27.001 5500 6.000V S 2W 27.002 EVAL Is the CH[MEM] recording at the {right edge of chart +/-1div}? 28.001 5500 0.000V S 2W 28.002 MESS Is the chart recording at the same center line as the 28.003 EVAL previous center check +/- {1div (dead band)}? 29.001 MESS 29.002 DISP Using the keyboard on the side of the UUT... 29.002 DISP 29.002 DISP Set the CHANNEL NUMBER to [MEM]. 29.002 DISP Set MEASUREMENT RANGE to 04. 29.002 DISP Set RECORDING RANGE to -20.00V to 20.00V. 29.003 5500 04 -20.00V 0.1% 0.02U 2W 30.001 5500 -20.00V S 2W 30.002 EVAL Is the CH[MEM] recording at the {left edge of chart +/-1div}? 31.001 5500 04 0.00V 0.1% 0.02U 2W 32.001 5500 0.00V S 2W 32.002 EVAL Is the CH[MEM] recording at the {center of chart +/-1div}? 33.001 5500 04 20.00V 0.1% 0.02U 2W 34.001 5500 20.00V S 2W 34.002 EVAL Is the CH[MEM] recording at the {right edge of chart +/-1div}? 35.001 5500 0.00V S 2W 35.002 MESS Is the chart recording at the same center line as the 35.003 EVAL previous center check +/- {1div (dead band)}? 36.001 MESS 36.002 DISP Using the keyboard on the side of the UUT... 36.002 DISP 36.002 DISP Set the CHANNEL NUMBER to [MEM]. 36.002 DISP Set MEASUREMENT RANGE to 05. 36.002 DISP Set RECORDING RANGE to -50.00V to 50.00V. 36.003 5500 05 -50.00V 0.1% 0.02U 2W 37.001 5500 -50.00V S 2W 37.002 EVAL Is the CH[MEM] recording at the {left edge of chart +/-1div}? 38.001 5500 05 0.00V 0.1% 0.02U 2W 39.001 5500 0.00V S 2W 39.002 EVAL Is the CH[MEM] recording at the {center of chart +/-1div}? 40.001 5500 05 50.00V 0.1% 0.02U 2W 41.001 5500 50.00V S 2W 41.002 EVAL Is the CH[MEM] recording at the {right edge of chart +/-1div}? 42.001 5500 0.00V S 2W 42.002 MESS Is the chart recording at the same center line as the 42.003 EVAL previous center check +/- {1div (dead band)}? 43.001 MESS 43.002 END